Operation of gas proportional scintillation counters in a low charge multiplica...

Author(s): Borges, F. I. G. M. cv logo 1 ; Santos, J. M. F. dos cv logo 2 ; Dias, T. H. V. T. cv logo 3 ; Santos, F. P. cv logo 4 ; Rachinhas, P. J. B. M. cv logo 5 ; Conde, C. A. N. cv logo 6

Date: 1999

Persistent ID: http://hdl.handle.net/10316/4555

Origin: Estudo Geral - Universidade de Coimbra


Description
The operation of a xenon filled gas proportional scintillation counter under low charge multiplication gains is discussed. It is shown that the best energy resolution for X-rays is obtained for a reduced electric field, E/p, of about 8 V cm-1 Torr-1, leading to a charge gain (for inter-grid distance of 5 mm) of about 1.05, as calculated by a Monte Carlo simulation method. A discussion is presented concerning the dependence of the E/p values for better detector performance, on the inter-grid distance. http://www.sciencedirect.com/science/article/B6TJM-3VR1CVW-29/1/cf66fb6eeb6e5fdc4e2c2bf6672578ae