Results for the energy resolution of a uniform-field gas proportional-scintillation counter are presented for the range 0.1-3 keV. A comparison with results from the literature shows that the detector is most suitable for this energy range. Examples of x-ray fluorescence pulse-height distributions from geological and industrial samples are presented. Copyright © 2004 John Wiley & Sons, Ltd. ; http://dx.doi.o...
A review on gas proportional scintillation counters (GPSCs) is presented. Recent achievements towards the portability of simple, inexpensive and compact GPSCs are discussed. Compensation of solid angle effects with the curved grid technique can be used to produce non-focused GPSCs with medium-sized radiation windows, at least up to 80% of the photosensor active diameter, without degradation of detector performa...
The operation of a xenon filled gas proportional scintillation counter under low charge multiplication gains is discussed. It is shown that the best energy resolution for X-rays is obtained for a reduced electric field, E/p, of about 8 V cm-1 Torr-1, leading to a charge gain (for inter-grid distance of 5 mm) of about 1.05, as calculated by a Monte Carlo simulation method. A discussion is presented concerning th...
The gas multiplication factor M and energy resolution R of xenon gas cylindrical proportional counters are investigated experimentally and calculated theoretically using a Monte Carlo technique to simulate the growth of single-electron-initiated avalanches. A good agreement is found between calculated and experimental data. The experimental and the calculated results are presented as a function of the reduced v...
The results of an experimental determination of the pulse-height distortion in soft X-ray spectra in gaseous xenon detectors are presented. The study confirms previous Monte Carlo simulation results and emphasizes the importance of the electric field in the drift region in reducing spectral degradation while the effect of a strong magnetic field in the same region is shown to be negligible ; JNICT (Junta Nac...