X-ray spectrometry with Peltier-cooled large area avalanche photodiodes

Autor(es): Fernandes, L. M. P. cv logo 1 ; Lopes, J. A. M. cv logo 2 ; Santos, J. M. F. dos cv logo 3 ; Conde, C. A. N. cv logo 4

Data: 2004

Identificador Persistente: http://hdl.handle.net/10316/4458

Origem: Estudo Geral - Universidade de Coimbra

Assunto(s): Avalanche photodiode; X-ray spectrometry; Peltier-cooling


Descrição
Performance characteristics of the response of a Peltier-cooled large-area avalanche photodiode are investigated. Detector gain, energy linearity, energy resolution and minimum detectable energy are studied at different operation temperatures. Detector energy resolution and lowest detectable X-ray energy present a strong improvement as the operation temperature is reduced from 25 to 15 °C and slower improvements are achieved for temperatures below 10 °C. http://www.sciencedirect.com/science/article/B6TJN-48W8TPB-3/1/f2e5cc2f92ebf5121b66551085272dd4