Accuracy and repeatability of a new portable ultrasound pachymeter

Author(s): Queirós, A. cv logo 1 ; González-Méijome, José Manuel cv logo 2 ; Fernandes, Paulo Rodrigues cv logo 3 ; Jorge, Jorge cv logo 4 ; Almeida, José B. cv logo 5 ; Parafita, Manuel A. cv logo 6

Date: 2007

Persistent ID: http://hdl.handle.net/1822/6199

Origin: RepositóriUM - Universidade do Minho

Subject(s): Pachymeter comparison; Portable ultrasound pachymetry; Corneal thickness


Description
To assess the accuracy and repeatability of central corneal thickness (CCT) measurements taken with a new portable ultrasound (US) pachymeter. Methods: Central thickness measurements were taken with a portable pachymeter (SP-100 Handy;Tomey, Nagoya, Japan) and a conventional US pachymeter (Nidek UP-1000; Nidek Technologies, Gamagori, Japan) from 57 right corneas of 57 young adults (19 males, 38 females) aged 18–44 years (mean ± S.D., 22.95 ± 3.92). Three repeated measures were obtained and then compared to obtain the repeatability of each instrument and the agreement between the pachymeters. The three readings taken with the portable pachymeter were compared against each other in order to evaluate intra-session repeatability and bias of each individual measurement with respect to the mean of three. Results: Mean values of CCT were 537 ± 35 micron for conventional and 534 ± 35 micron for the new portable pachymeter. A high agreement was found between the two instruments mean difference = 2.58 micron; 95% CI 1.41–3.75 micron) with only two eyes presenting differences larger than ± 8.6 micron which represents 95% CI in the Bland-Altman plots which represents 1.6% of the mean CCT. The first reading taken showed the highest agreement with the mean value for the portable pachymeter. Conclusions: The instrument tested in this study is able to take reliable measurements of corneal thickness even if a single reading is considered. Intra-session repeatability was very high, as was also the agreement between the average of three readings taken with the two US pachymeters.
Document Type Article
Language English
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Related documents



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento EU