Retinoscopy/autorefraction : which is the best starting point for a noncyclople...

Autor(es): Jorge, Jorge cv logo 1 ; Almeida, José B. cv logo 2 ; Parafita, Manuel A. cv logo 3 ; Queirós, A. cv logo 4

Data: 2005

Identificador Persistente: http://hdl.handle.net/1822/5799

Origem: RepositóriUM - Universidade do Minho

Assunto(s): Refraction; Refractive error; Accuracy; Automated refractor; Retinoscopy; Subjective refraction; Orthogonal functions; Astigmatism


Descrição
Purpose: The aim of this study was to estimate the agreement between an autorefractor (Nidek ARK 700A) and retinoscopy with subjective refraction. Methods: Measurements of autorefraction obtained with the ARK700A and retinoscopy were performed on 192 right eyes from 192 healthy young adults and compared with subjective refraction. These measurements were performed without cycloplegia. The age range was 18-34 years, with a mean value of 21.6 years and a standard deviation (SD) of 2.66 years. Results: A comparison of the autorefractor and subjective refraction results shows that; (1) for the mean spherical equivalent (M) the autorefractor yields more negative values (–0.44 ± 0.54 D, p = 0.000); (2) for the Jackson cross-cylinder at axis 0º (J0), the autorefractor yields more positive values than the subjective ones (–0.05 ± 0.13 D, p = 0.000); (3) and for the Jackson cross-cylinder at axis 45º (J45), the autorefractor results are more negative (–0.02 ± 0.09 D, p = 0.019). The differences found for each component M, J0 e J45 are statistically significant. By comparing retinoscopy with the subjective exam, there are no statistically significant differences found for the M component, (–0.02 ± 0.33 D, p = 0.304). For the J0 and J45 components, the differences are statistically significant (–0.07 ± 0.10 D, p = 0.000;–0.01 ± 0.08 D, p = 0.008). Conclusions: the present results confirm that when performed by an experienced clinician, retinoscopy is more accurate than automatic refraction giving a better starting point to non-cycloplegic refraction.
Tipo de Documento Artigo
Idioma Inglês
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Documentos Relacionados



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento União Europeia