Detalhes do Documento

An RPC-PET prototype with high spatial resolution

Autor(es): Blanco, A. cv logo 1 ; Carolino, N. cv logo 2 ; Correia, C. M. B. A. cv logo 3 ; Marques, R. Ferreira cv logo 4 ; Fonte, P. cv logo 5 ; González-Díaz, D. cv logo 6 ; Lindote, A. cv logo 7 ; Lopes, M. I. cv logo 8 ; Macedo, M. P. cv logo 9 ; Policarpo, A. cv logo 10

Data: 2004

Identificador Persistente: http://hdl.handle.net/10316/4436

Origem: Estudo Geral - Universidade de Coimbra

Assunto(s): Gaseous detectors; Photodetectors; Resistive plate chambers (RPC); Positron emission tomography (PET); Small animal imaging


Descrição
A small positron emission tomography system, based on the timing RPC technology has been built and tested. This first prototype is aimed at validating the expectations, derived from simulations, of a very high spatial resolution, which could be of value for the imaging of small animals. The system is composed of two counting heads, able to measure the photon interaction point in two dimensions, the transaxial dimension and the Depth of Interaction. Each head is composed of 16 independent stacked RPCs made from copper and glass (anode) electrodes. Point-like 22Na sources were so far reconstructed, using the standard filtered back-projection algorithm, with a spatial accuracy of 0.6 mm FWHM, free of parallax error. http://www.sciencedirect.com/science/article/B6TJM-4CXB0P8-4/1/5e71356440cf220c1ce79dbe820b564d
Tipo de Documento Artigo
Idioma Inglês
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Documentos Relacionados



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento União Europeia