Pulse-height-spectrum distortion in xenon gaseous detectors for soft X-rays: ex...

Autor(es): Santos, F. P. cv logo 1 ; Santos, J. M. F. dos cv logo 2 ; Dias, T. H. V. T. cv logo 3 ; Conde, C. A. N. cv logo 4

Data: 1995

Identificador Persistente: http://hdl.handle.net/10316/12895

Origem: Estudo Geral - Universidade de Coimbra


Descrição
The results of an experimental determination of the pulse-height distortion in soft X-ray spectra in gaseous xenon detectors are presented. The study confirms previous Monte Carlo simulation results and emphasizes the importance of the electric field in the drift region in reducing spectral degradation while the effect of a strong magnetic field in the same region is shown to be negligible JNICT (Junta Nacional de Investigação Científica e Tecnológica) Projecto STRDA/P/CEN/422/92