Document details

Sterilization of an Electronical Medical Device

Author(s): Reis, Joana cv logo 1 ; Cabo Verde, Sandra cv logo 2 ; Capela Silva, Fernando cv logo 3 ; Potes, José cv logo 4 ; Melo, Rita cv logo 5 ; Nunes, Inês cv logo 6 ; Marcos, Helena cv logo 7 ; Silva, Telma cv logo 8 ; Botelho, Maria Luísa cv logo 9

Date: 2010

Persistent ID: http://hdl.handle.net/10174/2760

Origin: Repositório Científico da Universidade de Évora

Subject(s): ionizing radiation; ultraviolet radiation; programmable memories


Description
Radiosterilization was applied to a medical device with a “programmable memory” to allow in vivo implantation. Irradiation on a Cobalt-60 facility at 25 kGy at a dose rate of 2kGy/h corrupted the memory. Therefore an alternative sterilization method using UV was developed and validated based on ISO 11737-1 and ISO 14937. These procedures may be useful and effective for research purposes when only a small number of items might be involved but applicability at an industrial scale is unlikely.
Document Type Article
Language English
Editor(s) NAARRI International Conference – NIC 2010
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Related documents



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento EU