Detalhes do Documento

Electroluminescence yield in xenon gas detectors

Autor(es): C.M.B. Monteiro cv logo 1 ; L.M.P. Fernandes cv logo 2 ; J.A.M. Lopes cv logo 3 ; L.C.C. Coelho cv logo 4 ; J.F.C.A. Veloso cv logo 5 ; J.M.F.dos Santos cv logo 6 ; K. Giboni cv logo 7 ; E. Aprile cv logo 8

Data: 2012

Origem: Electrónica e Telecomunicações

Assunto(s): electroluminescence yield; xenon gas; scintillation; xenon gas detectors; electroluminescence yield; xenon gas; scintillation; xenon gas detectors


Descrição
The electroluminescence yield was studied for xenon gas as a function of the electric field in the scintillationregion, for room temperature using a gas proportional scintillation counter. A large area avalanche photodiode was used for the readout of the VUV secondary scintillation produced in the gas, together with the 5.9-keV x-rays directly absorbed in the photodiode. Using the latter as a reference for the number of charge carriers produced by the scintillation pulse, it was possible to determine the number of VUV photons impinging the photodiode. This way, a scintillation amplification parameter of 140 photons/kV was obtained. The attained results are in good agreement with those predicted, for room temperature, by Monte Carlo simulation and Boltzmann calculations, as well as with those obtained for saturated xenon vapour, at cryogenic temperatures, and are about a factor of two higher than former results measured at room temperature. The electroluminescence yield was studied for xenon gas as a function of the electric field in the scintillationregion, for room temperature using a gas proportional scintillation counter. A large area avalanche photodiode was used for the readout of the VUV secondary scintillation produced in the gas, together with the 5.9-keV x-rays directly absorbed in the photodiode. Using the latter as a reference for the number of charge carriers produced by the scintillation pulse, it was possible to determine the number of VUV photons impinging the photodiode. This way, a scintillation amplification parameter of 140 photons/kV was obtained. The attained results are in good agreement with those predicted, for room temperature, by Monte Carlo simulation and Boltzmann calculations, as well as with those obtained for saturated xenon vapour, at cryogenic temperatures, and are about a factor of two higher than former results measured at room temperature.
Tipo de Documento Documento de conferência
Idioma Inglês
Editor(es) ;
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Documentos Relacionados



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento União Europeia