A novel continuous small-scale reactor based on the oscillatory flow technology (Harvey
et al 2001) is being developed for application to specialist chemical manufacture and high
throughput continuous screening (Harvey et al 2003). This novel reactor is able to perform
continuous multiphase reactions, including those involving the suspension of catalyst beads.
Extensive potential use in chemistry, biological and pharmaceutical laboratories is envisaged.
Optimum operation conditions for applications in the bioengineering field depend on at
least four parameters: 1) fluid mixing, 2) residence time characteristics, 3) particle suspension
and 4) (oxygen) mass transfer rates. This work particularly concerns the establishment of the
operation conditions in relation to oxygen mass transfer rates, KL.a. Furthermore, KL.a values
are correlated with the fluid mixing, axial dispersion coefficients and finally with the fluid
mechanics observed experimentally by particle image velocimetry (PIV) technique and
numerically simulated using Fluent® software.
The screening reactor is composed of a 35 cm length glass jacketed tubes (Figure 1),
provided with smooth periodic constrictions (SPCs). The internal diameter of the tube is 4.4
mm. The diameter in the constricted zone (the baffle internal diameter) is about 1.6 mm,
representing an 87 % reduction in the cross-sectional area. Mixing intensity is controlled by
setting up the frequency and the amplitude (centre-to-peak) of the fluid oscillations. Typical
oscillation frequencies and amplitudes are from 0 to 20 Hz and from 0 to 4 mm, respectively.
A screening arrangement based on some SPC tubes (say 10 to 20) placed at different
configurations (serial or parallel) makes this novel reactor suitable for parallel processing
and/or for sequential reactions procedures.
State-of-the-art fibre-optical technology was used for on-line monitoring of the oxygen
concentration inside the screening reactor by using a special fibre optical micro-probe. The
working tip of the probe was dip-coated with a ruthenium complex immobilised in a sol-gel
matrix. This complex was excited to fluorescence by a blue led (≈ 470 nm output peak) and
the level of such fluorescence is inversely related with the concentration of the oxygen
through the Stern-Volmer equation (Wang et al., 1999). Continuous fluorescence levels are
accurately measured by an UV/VIS/NIR multi-channel spectrometer.
Numerical simulations by the computational fluid dynamics (CFD) technique, using
Fluent ® software, permitted us to conclude that either near plug flow or stirred tank
behaviour can be approached in a single SPC tube, by controlling the fluid oscillation
conditions, i.e. the oscillation frequency and/or amplitude. For oscillatory Reynolds numbers
(Reo) between 10 and 100, the formation of axisymmetric vortex rings leads to a good radial
mixing of the fluid and to low axial dispersions, which suggests that a performance near a
plug flow reactor is achieved. For Reo above 100 the high intensity and asymmetry of the
vortex rings leads to an increase of the axial dispersion and the fluid behaviour approach that
of a stirred tank. All these results were experimentally validated by PIV observations.
High radial rates of flow exchange were numerically simulated and experimentally
observed during a complete oscillation cycle, coupled with a high degree of velocity
gradients. Thus, enhancements of heat and mass transfer rates are expectable within this novel
screening reactor. Further, an increase of bubbles breakage is also expected (leading to a
decrease of bubble diameter, i.e. an increase of the specific bubble area, a, and also of the gas
hold-up), conducting to a significant improvement of oxygen mass transfer. This will be
demonstrated with the present work.