Document details

An array of fabry-perot optical-channels for biological fluids analysis

Author(s): Minas, Graça cv logo 1 ; Ribeiro, J. C. cv logo 2 ; Martins, Júlio S. cv logo 3 ; Wolffenbuttel, R. F. cv logo 4 ; Correia, J. H. cv logo 5

Date: 2004

Persistent ID: http://hdl.handle.net/1822/1339

Origin: RepositóriUM - Universidade do Minho

Subject(s): Biosystem; Optical filter array; Fabry-Perot


Description
This paper describes a biosystem (biological system) used to measure the concentration of biochemical substances in urine, serum, plasma or cerebrospinal fluid. Rather than just one channel, it comprises 16 optical-channels that enable the measurement of the concentration of 16 different biochemical substances. An array of 16 optical filters based on Fabry-Perot thin-films optical resonators has been designed. Each optical-channel is sensitive in a single wavelength with a full-width-half-maximum (FWHM) of 7 nm. The filter fabrication requires only four masks, used with different etch time. A commercially available band-pass optical filter with a band-pass wavelength in 450–650 nm is used. The biosystem requires only a white light source for illumination due the use of selective optical filters.
Document Type Article
Language English
delicious logo  facebook logo  linkedin logo  twitter logo 