Detalhes do Documento

Multi-syringe flow injection system with in-line microwave digestion for the d...

Autor(es): Almeida, M. Inês G.S. cv logo 1 ; Segundo, Marcela A. cv logo 2 ; Lima, José L.F.C. cv logo 3 ; Rangel, António O.S.S. cv logo 4

Data: 2004

Identificador Persistente: http://hdl.handle.net/10400.14/3515

Origem: Veritati - Repositório Institucional da Universidade Católica Portuguesa

Assunto(s): Multi-syringe flow analysis; Total phosphorus; Microwave; In-line digestion; Waste waters


Descrição
A multi-syringe system for spectrophotometric determination of total phosphorus involving in-line digestion is proposed. Sample and digestion solution were dispensed and directed towards a digestion vessel located inside a domestic microwave oven (MWO) where sample digestion took place. Afterwards, the digested sample was merged with the necessary reagents for the colorimetric determination based on the molybdenum blue method. Several digestion conditions were studied regarding composition of digestion solution, digestion time and power set on theMWO. The system was applied to waste water samples and results shown a good agreement with the reference method. Repeatable results (R.S.D. < 2.41%) and determination frequency of 12 h−1 were obtained.
Tipo de Documento Artigo
Idioma Inglês
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Documentos Relacionados



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento União Europeia