Determination of nitrate and nitrite in dairy samples by sequential injection u...

Autor(es): Lima, M.J. Reis cv logo 1 ; Fernandes, Sílvia M.V. cv logo 2 ; Rangel, António O.S.S. cv logo 3

Data: 2006

Identificador Persistente: http://hdl.handle.net/10400.14/2852

Origem: Veritati - Repositório Institucional da Universidade Católica Portuguesa

Assunto(s): Sequential injection; Copperized cadmium column; Nitrate; Nitrite; Dairy samples


Descrição
A sequential injection analysis (SIA) system for the spectrophotometric determination of nitrate in dairy samples was developed. A test portion was aspirated into a carrier solution containing ethylendiaminetetracetic acid (EDTA) and ammonium buffer, which flowed into a copperized cadmium reduction column installed in-line for the determination of the nitrate plus nitrite contents of samples. For the nitrite determination, another test portion of sample was aspirated and directly sent to the detector without reduction. Nitrate content was calculated from the difference between nitrate plus nitrite (expressed as nitrite) and nitrite content. The spectrophotometric determination is based on the Griess reaction. The proposed method was used to test several dairy samples (ultrapasteurized milk with 1.7% milk fat, whey, raw bovine milk and several cheese varieties). Results were statistically in good agreement with those provided by the reference procedure, with a detection limit of 0.15 mgL 1. A sampling rate of 21 determinations per hour can be achieved with this procedure.
Tipo de Documento Artigo
Idioma Inglês
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Documentos Relacionados



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento União Europeia