Detalhes do Documento

Evaluation of exposure index (IgM) in orthopaedic radiography

Autor(es): Lança, Luís cv logo 1 ; Silva, Augusto cv logo 2

Data: 2008

Identificador Persistente: http://hdl.handle.net/10400.21/991

Origem: Repositório Científico do Instituto Politécnico de Lisboa

Assunto(s): Radiologia; Radiografia ortopédica; Dosimetria; Radiology; Orthopaedic radiography; Dosimetry


Descrição
The exposure index (lgM) obtained from a radiographic image may be a useful feedback indicator to the radiographer about the appropriate exposure level in routine clinical practice. This study aims to evaluate lgM in orthopaedic radiography performed in the standard clinical environment. We analysed the lgM of 267 exposures performed with an AGFA CR system. The mean value of lgM in our sample is 2.14. A significant difference (P=0.000<0.05) from 1.96 lgM reference is shown. Data show that 72% of exposures are above the 1.96 lgM and 42% are above the limit of 2.26. Median values of lgM are above 1.96 and below 2.26 for Speed class (SC) 200 (2.16) and SC400 (2.13). The interquartile range is lower in SC400 than in SC200. Data seem to indicate that lgM values are above the manufacturer’s reference of 1.96. Departmental exposure charts should be optimised to reduce the dose given to patients.
Tipo de Documento Artigo
Idioma Inglês
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Documentos Relacionados



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento União Europeia