Detalhes do Documento

Novel polymer-modified electrodes for batch injection sensors and application t...

Autor(es): Brett, Christopher M. A. cv logo 1 ; Fungaro, Denise A. cv logo 2 ; Morgado, José M. cv logo 3 ; Gil, M. Helena cv logo 4

Data: 1999

Identificador Persistente: http://hdl.handle.net/10316/5262

Origem: Estudo Geral - Universidade de Coimbra

Assunto(s): Anodic stripping voltammetry; Polymer-coated electrodes; Batch injection analysis; Environmental electroanalysis


Descrição
Various polymer coatings have been investigated for the protection of mercury thin-film electrodes in the square wave anodic stripping voltammetry of environmental samples using batch injection analysis, with injection of untreated samples of volume 50 [mu]l directly over the sensing electrode. Polymer coatings studied include those with controlled porosity, and cation-exchange polymers based on sulphonated polymers. Of the polymers tested, films of ca. 1 [mu]m thickness made from Nafion® mixed with 5% poly(vinyl sulphonic acid) were found to give the best results in tests with the model surfactants Triton-X-100 detergent, sodium dodecyl sulphate and protein standard. The validity of the approach is demonstrated by application to real samples. http://www.sciencedirect.com/science/article/B6TGB-3WS69WY-1G/1/482035fd36bbbdaace32d2213fe93529
Tipo de Documento Artigo
Idioma Inglês
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Documentos Relacionados



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento União Europeia