Detalhes do Documento

Simulation and experimental results for the detection of conversion electrons w...

Autor(es): Rachinhas, P. J. B. M. cv logo 1 ; Simões, P. C. P. S. cv logo 2 ; Lopes, J. A. M. cv logo 3 ; Dias, T. H. V. T. cv logo 4 ; Morgado, R. E. cv logo 5 ; Santos, J. M. F. dos cv logo 6 ; Stauffer, A. D. cv logo 7 ; Conde, C. A. N. cv logo 8

Data: 2000

Identificador Persistente: http://hdl.handle.net/10316/4538

Origem: Estudo Geral - Universidade de Coimbra

Assunto(s): Gas scintillation detectors; Conversion electrons; Monte Carlo simulation; Digital pulse processing


Descrição
The application of gas proportional scintillation counters (GPSC) to the detection and identification of conversion electrons in the medium energy range is investigated. Experimental and Monte Carlo results are presented for the response of a xenon GPSC, filled at atmospheric pressure, to the decay of a 109Cd source. This source emits 88.0 keV [gamma]-rays, eL=84.6 keV and eK=62.5 keV conversion electrons, as well as fluorescence X-rays and Auger electrons. Good agreement is found between the measured and the calculated energy spectra. The response to higher-energy electrons is investigated by Monte Carlo simulation, by considering a hypothetical GPSC filled with xenon at 10 atm and doped with the 133mXe metastable isotope. The calculated energy spectra for the absorption of the 133mXe 233.2 keV [gamma]-rays, eK=198.6 keV and eL=228.4 keV conversion electrons, as well as fluorescence X-rays and Auger electrons, are presented and discussed. http://www.sciencedirect.com/science/article/B6TJM-3YS91CB-J/1/4b48a2a89f10ac68be0e04c4537a9d54
Tipo de Documento Artigo
Idioma Inglês
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Documentos Relacionados



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento União Europeia