Detalhes do Documento

A technique for the absolute measurement of the W-value for X-rays in counting ...

Autor(es): Vinagre, F. L. R. cv logo 1 ; Conde, C. A. N. cv logo 2

Data: 2000

Identificador Persistente: http://hdl.handle.net/10316/4527

Origem: Estudo Geral - Universidade de Coimbra

Assunto(s): W-value; Ionization yield; Radiation physics; Dosimetry; X-ray detectors


Descrição
A technique was developed for the absolute measurement of the W-value (the mean energy for the production of an electron-ion pair) for low-energy X-rays in a wide range of gases at atmospheric pressures, with a standard uncertainty better than 1%. This technique is based on the absolute measurement of the primary ionization charge produced by X-ray photons from a constant intensity monoenergetic X-ray source, e.g. a long lifetime radioactive source. The ionization charge is calibrated by the number of X-ray photons absorbed in the gas, counted with a photon detector. For this purpose, a hybrid detector system was tested and its use in W-value measurements was investigated. The technique was applied to pure xenon at 825 Torr with 5.9 keV X-rays and a W-value of 21.61-0.10+0.14 eV was obtained for a 68% confidence level. The required corrections and the different factors contributing to the accuracy of the results are discussed. The advantages and limitations of this technique are explored and future developments are discussed. http://www.sciencedirect.com/science/article/B6TJM-40WDR8N-N/1/6c1c7baa3952be96f48aa99c589e7a18
Tipo de Documento Artigo
Idioma Inglês
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Documentos Relacionados



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento União Europeia