The gas proportional scintillation counter/microstrip gas chamber hybrid detector

Author(s): Freitas, D. S. A. P. cv logo 1 ; Veloso, J. F. C. A. cv logo 2 ; Monteiro, C. M. B. cv logo 3 ; Santos, J. M. F. dos cv logo 4 ; Conde, C. A. N. cv logo 5

Date: 2003

Persistent ID: http://hdl.handle.net/10316/4472

Origin: Estudo Geral - Universidade de Coimbra

Subject(s): Gas scintillation counter; CsI photocathode


Description
The gas proportional scintillation counter/microstrip gas chamber (GPSC/MSGC) hybrid detector for X-ray spectrometry is described. The detector uses a CsI-coated microstrip plate placed in direct contact with the gas-filling as the photosensor readout for the GPSC scintillation substituting for the photomultiplier tube (PMT). Usable photosensor maximum gain is limited by optical positive feedback due to the additional scintillation produced in the electron avalanche process at the MSP anodes, in the absence of quenching. A low-photoelectron collection efficiency is achieved in the gas atmosphere, resulting in a scintillation conversion efficiency that is about a factor of 5 lower than that achieved with PMT-based GPSCs. However, energy resolutions of 11% for 5.9 keV X-rays are achieved with this detector. http://www.sciencedirect.com/science/article/B6TJM-486BN6Y-M/1/4da741da1d82059bc939bc52156b6569
Document Type Article
Language English