Document details

MHSP operation in pure xenon

Author(s): Luz, H. Natal da cv logo 1 ; Veloso, J. F. C. A. cv logo 2 ; Amaro, F. D. cv logo 3 ; Ferreira, L. F. Requicha cv logo 4 ; Santos, J. M. F. dos cv logo 5 ; Breskin, A cv logo 6 ; Chechik, R. cv logo 7

Date: 2005

Persistent ID: http://hdl.handle.net/10316/4417

Origin: Estudo Geral - Universidade de Coimbra

Subject(s): Gas electron multiplier


Description
We present latest results of a micro-hole and strip plate (MHSP) electron multiplier operating in pure xenon at atmospheric pressure. We report on avalanche gain of ~2×104, stable within ~3% after an accumulated charge of 4 [mu]C/mm2, when irradiated with 0.5 kHz/mm2 5.9 keV photons. No photon-induced secondary effects or discharges have been observed at this gain. An energy resolution of 14% was obtained with 5.9 keV X-rays. http://www.sciencedirect.com/science/article/B6TJM-4GPVP8N-4/1/8f46f4fef69febd247768213084b087b