Spatial resolution on a small animal RPC-PET prototype operating under magnetic...

Author(s): Blanco, A. cv logo 1 ; Carolino, N. cv logo 2 ; Correia, C. M. B. A. cv logo 3 ; Fazendeiro, L. cv logo 4 ; Ferreira, Nuno C. cv logo 5 ; Marques, M. F. Ferreira cv logo 6 ; Marques, R. Ferreira cv logo 7 ; Fonte, P. cv logo 8 ; Gil, C. cv logo 9 ; Macedo, M. P. cv logo 10

Date: 2006

Persistent ID: http://hdl.handle.net/10316/4399

Origin: Estudo Geral - Universidade de Coimbra

Subject(s): Gaseous detectors; magnetic field


Description
It has been demonstrated in previous work that the RPC-PET technology is able to deliver radioisotope image resolutions approaching the physical limits of the PET principle. Here we study, by simulation, the effect of the magnetic field on the positron range to evaluate whether the spatial resolution of the RPC-PET could be improved by using an intense magnetic field. Six positron emitters of interest to small animal PET imaging (18F, 11C, 15O, 68G, 62Cu and 86Y) are considered. Results suggest that a three-fold improvement on the spatial resolution may be obtained under a magnetic field of 10 T for the higher energy radioisotopes like 86Y or 62Cu, and by about 20% for the lower energy ones, like 18F or 11C. http://www.sciencedirect.com/science/article/B6TVD-4KV4MR8-15/1/efa4356ba16a85fd7740937bd0e672a9