Document details

Pulse-height-spectrum distortion in xenon gaseous detectors for soft X-rays: ex...

Author(s): Santos, F. P. cv logo 1 ; Santos, J. M. F. dos cv logo 2 ; Dias, T. H. V. T. cv logo 3 ; Conde, C. A. N. cv logo 4

Date: 1995

Persistent ID: http://hdl.handle.net/10316/12895

Origin: Estudo Geral - Universidade de Coimbra


Description
The results of an experimental determination of the pulse-height distortion in soft X-ray spectra in gaseous xenon detectors are presented. The study confirms previous Monte Carlo simulation results and emphasizes the importance of the electric field in the drift region in reducing spectral degradation while the effect of a strong magnetic field in the same region is shown to be negligible JNICT (Junta Nacional de Investigação Científica e Tecnológica) Projecto STRDA/P/CEN/422/92
Document Type Article
Language English
delicious logo  facebook logo  linkedin logo  twitter logo 
degois logo
mendeley logo

Related documents



    Financiadores do RCAAP

Fundação para a Ciência e a Tecnologia Universidade do Minho   Governo Português Ministério da Educação e Ciência Programa Operacional da Sociedade do Conhecimento EU