Assessment of 3D scanners for modeling Pectus Carinatum corrective bar

Autor(es): Moreira, António H. J. cv logo 1 ; Fonseca, João Gomes cv logo 2 ; Rodrigues, Pedro L. cv logo 3 ; Fonseca, Jaime C. cv logo 4 ; Pinho, A.C.M. cv logo 5 ; Correia-Pinto, Jorge cv logo 6 ; F. Rodrigues, Nuno cv logo 7 ; Vilaça, João L. cv logo 8

Data: 2013

Identificador Persistente: http://hdl.handle.net/11110/509

Origem: CiencIPCA

Assunto(s): 3D scanner; Kinect; FastSCAN; Pectus Carinatum; Surface mesh; 3D reconstruction


Descrição
Pectus Carinatum (PC) is a chest deformity consisting on the anterior protrusion of the sternum and adjacent costal cartilages. Non-operative corrections, such as the orthotic compression brace, require previous information of the patient chest surface, to improve the overall brace fit. This paper focuses on the validation of the Kinect scanner for the modelling of an orthotic compression brace for the correction of Pectus Carinatum. To this extent, a phantom chest wall surface was acquired using two scanner systems – Kinect and Polhemus FastSCAN – and compared through CT. The results show a RMS error of 3.25mm between the CT data and the surface mesh from the Kinect sensor and 1.5mm from the FastSCAN sensor
Tipo de Documento Artigo