Document details

Development and evaluation of a normal phase liquid chromatographic method for ...

Author(s): Amaral, J.S. cv logo 1 ; Casal, Susana cv logo 2 ; Seabra, R.M. cv logo 3 ; Oliveira, M.B.P.P. cv logo 4

Date: 2005

Persistent ID: http://hdl.handle.net/10198/632

Origin: Biblioteca Digital do IPB

Subject(s): Tocopherol; Tocotrienol; Normal phase liquid chromatography; Walnuts


Description
A high performance liquid chromatographic (HPLC) method for the determination of tocopherols and tocotrienols in walnut samples is described. The compounds were extracted with n-hexane, using a simple solid-liquid extraction procedure. Tocol was used as internal standard and BHT as anti-oxidant. The ehromatographic separation was achieved using an Inertsil 5 SI normal phase column operating with isocratic elution of n-hexane/1,4-dioxane (96.5:3.5, v/v), at a flow rate of 0.7 mL/min. The effluent was monitored by a series arrangement of a diodearray detector followed by a fluorescence detector. The detection limits were low, between 0.037 and 0.266 (jig/mL. The method was precise (% CV less than2.8%), accurate (% CV less than 5.6%), and, as a general rule, the recovery values were high (mean values ranging from 93.4% to 104%).